Compatibility with Luminex 100
Version 1.7

Certain settings in the Luminex 100 Version 1.7 software need to
have particular values for your assay to be compatible with the
Luminex 100 IS system. See Table 11.

Note:

You cannot have Luminex 100 Version 1.7 and Luminex 100

IS loaded on the same system. Record your gate values on the
Luminex 100 Version 1.7 system and then enter the information into
templates on the Luminex 100 IS system.

‹

To set the 1.7 Luminex bead map compatible to Luminex 100 IS

Version 2.2:

1. Select the Luminex 100 Version 1.7 icon on your desktop. The

application loads and displays the opening screen.

2. Click Options on the toolbar. On the User tab of the Options

dialog box, make sure that the Default Bead Map is set to 100
regions
. If it is not, click Browse and select the 100 regions bead
map. Click OK. See Figure 64.

A

Table 11: Value Settings for the Luminex 100

Default bead map:

100 regions

Sample size:

between 20 and 200

µL

Events per bead:

recommend 100

Flow rate:

Fast