beautypg.com

Luminex 100 IS Version 2.2 User Manual

Page 49

background image

x

MAP

Technology

Create Templates

PN 89-00002-00-032 Rev. A

43

If desired, you can use the Apply Curve Fit button to apply the
same formula, that you selected on the first line, to all other tests.

15. Click Next to move to the next Template Setup Wizard dialog

box. See Figure 28.

Figure 28. Template Setup Wizard Dialog Box—Select

Template Commands

16. Select commands from the Available Commands list and move

them to the Template Commands list using the single arrow
pointing toward the template command button. A shortcut to
repeat the same commands is to enter the number in the Repeat
field and click the arrow next to the field. Figure 28 shows four
Acquire commands were entered with Repeat.

These commands become part of the template and are always
performed in the order specified for any batch that uses this
template. Each well command is automatically assigned to the
next available well. Laboratory personnel can select the starting
location on the plate, but can not change these commands or the
order that they are performed. Remember that if you have more
than one command that involves the reservoir, those commands
must use the same solution.