11. In the Min Beads field, enter the number of beads to collect for

each test. The Apply Min Beads button applies the same min
bead requirement that you entered on the first line to all other
tests.

12. Click Next to move to the next Template Setup Wizard dialog

box. See Figure 42.

Figure 42. Template Setup Wizard Dialog Box—Select

Template Commands

13. Select commands from the Available Commands list and move

them to the Template Commands list using the single arrow (>)
pointing towards the template command button.

These commands are then part of the template and are always
performed in the order specified for any batch that uses this
template. Each well command is automatically assigned to the
next available well. Laboratory personnel running the batch can
not change these commands. Remember that if you have more
than one command that involves the reservoir, those commands
must use the same solution.

For a complete list of command description, see Table 9 on
page 44.

14. After you enter all the template information, check it carefully.

When all the information is correct, click Save to save the
template, then click Close to exit.