13. In the Min Beads field, enter the number of beads to collect for

each test. The Apply Min Beads button applies the same min
bead requirement that you entered on the first line to all other
tests.

14. Click a Curve Fit field to see a list of formulas. Select one of the

curve fit formulas for each test. The available curve fit formulas
are listed in Table 8. For batches with replicate standards the
calculation of curve fit defaults to “Fit of All Standards”. If
desired, you can select curve fit to be drawn using “Mean of
Replicates”.

The algorithm for each formula appears on the selection list.

Figure 27. Template Setup Wizard—Select Curve Fit Formula

Table 8: Curve Fit Formulas

Linear: y=ax+b
Quadratic: y=a+b+c(x^2)
Cubic: y=a+bx+cx^2+dx^3
Logistic4p: y=a+b/(1+(x/c)^d)
Logistic5p: y=a+b/(1+(x/c)^d)^f
Power: y=a*(x^b)
Raw Formula: y=x
RiseToMax,3P: y=a-b*exp(cx)
RiseTo Max,2p: y=a-a*exp(bx)
Logarithmic: y=aInx+b
Exponential: y=a e^xb

Select a Curve Fit