When you replay a batch it labels unknown samples as Pa1, Pa2,
Pa3, and so on.

If you replay a batch containing replicates, replicate averaging will
not be calculated in data analysis.

A number of variables can affect the final test results. You may also
change the standards or controls processed with the batch or multi-
batch. These variables may effect your test results:

minimum number of events for acquisition

formula used to analyze the MFI values

standards or controls validation or invalidation

type of analysis (qualitative, quantitative, acquisition only, or
maintenance)

Reprocess

Samples Using

Replay Batch

You reprocess samples using Replay Batch from the Acquisition
Detail tab.

‹

To reprocess samples using Replay Batch:

1. Click the Acquisition Detail tab.

2. Click Replay Batch on the Acquisition Detail tab. A Browse for

Folder dialog box appears displaying the My Batches directory
(folder).

3. Select the desired batch under the My Batches folder and click

OK.

4. The Open Template dialog box appears. Select the desired

template and click Select.

5. The Run Batch tab becomes the active tab. You can monitor the

commands as they process. Switch to the Acquisition Detail tab
and monitor the data, histogram, and dot plot.

Analyze

Reprocessed Data

with Replay Batch

After running Replay Batch, you can analyze the data.

To analyze Replay Batch data:

1. Select Start Analysis on the toolbar.

2. At the Open Batch dialog box select the batch you want to

analyze and click Select. The most recent Replay Batch is the
last or has the highest ID number. See Figure 5.