Luminex 100 IS Version 2.2 User Manual
Page 58

Luminex
100
IS Developer Guide to
x
MAP Technology Version 2.2
x
MAP
Technology
52
PN 89-00002-00-032 Rev. A
15. Click Next to move to the next Template Setup Wizard dialog
box. See Figure 37.
Figure 37. Template Setup Wizard Dialog Box—Select
Template Commands
16. Select actions from the Available Commands list and move them
to the Template Commands list using the single arrow (>)
pointing towards the template command list.
These commands become part of the template and are always
performed in the order specified for any batch that uses this
Table 10: Algorithm Formulas
Qualitative
Analysis Method
Algorithm
Default
Values
Default
Ranges
Lum Qual Formula
0 to .9
.9 to 1.1
1.1 to 50
negative
low positive
high positive
Adv Qual Formula
0 to 20
20 to 40
40 to 60
60 to 100,000
negative
low positive
moderate
strong positive
Where: FI = fluorescent intensity
Ki = lot information for each test
FI
sample
FI
s
dard
tan
------------------------ Ki
×
FI
sample
FI
s
dard
tan
------------------------ Ki
Ч