Table 2 displays acceptable reporter fluorochromes wavelengths
excitation/emission.

Sample Size

Sampling volume or sample size can range from 20 to 200
microliters. Assays must have enough volume in the wells so that
after this sampling volume is taken up, some sample remains. Make
the volume in the wells at least 25 uL greater than the sample
volume. This amount may vary depending on the type of plate used.
After acquisition, the instrument washes the sample lines resulting in
ejection of approximately 160 microliters of sheath fluid back into
the well. Make sure that there is room to add this amount to the well
without overflowing the well and contaminating other wells.

Note:

The volume restrictions that the amount expelled back

function places on the assay design can be expounded by the
following formula.

Table 2: Reporter Fluorochromes

R-

Phycoerythrin

Alexa

532

Formula weight (Daltons)

240,000

470

Absorbance max (nm)

480, 546, 565

531

Extinction max (M

-1

cm

-1

)

1,960,000

83,800

Emission max (nm)

578

554

Quantum yield

0.82

0.8

[“Total well volume” (uL ) – “Sample uptake volume”(uL)] + 160(uL)

Where:

Total well volume

The starting sample volume of a well
before the unit samples for acquisition.

Sample uptake
volume

The volume the unit uptakes for
acquisition (program this in the template
as sample vol.)

160uL

The amount expelled back volume stated
in this section.

Maximum Well
volume plate

The maximum volume capacity of the
user’s selected 96 well microtiter plate.