Luminex 100 IS Version 2.2 User Manual
Page 50
Luminex
100
IS Developer Guide to
x
MAP Technology Version 2.2
x
MAP
Technology
44
PN 89-00002-00-032 Rev. A
Table 9: Available Commands
Command
Definition
Acquire
Acquires an unknown sample.
Acquire
Background
Acquires a background sample from a well.
The background results are automatically
subtracted from standards, controls, and
samples. If you are running more than one
background well, the software will subtract the
average of the background samples.
Acquire Control
Acquires the specified control from a well. If a
specified control is selected more
than once the software calculates the results
for each well independently. It then
reports an average of the control samples.
Acquire
Standard
Acquires the specified standard from a well. If
a specified standard is selected
more than once, the software calculates the
results for each well independently.
It then reports an average of the two samples.
Acquire Test
Specimen
Acquires a sample reading from a well. If
Acquire Test Specimen is selected more
than once, the software calculates the results
for each well independently. However,
samples having identical Sample IDs are
treated as replicates reporting an average
of the individual samples.
Backflush
Backflushes the system by reversing the flow
of sheath fluid through the cuvette.
Calibrate CAL1
Acquires calibration beads from a well to
calibrate the Classification and DD channels.
Calibrate CAL2
Acquires calibration beads from a well to
calibrate the Reporter channel.
Drain to
reservoir
Removes and refills sheath fluid from the
cuvette. Sheath fluid is expelled into the
reservoir.
Drain to well
Removes and refills sheath fluid from the
cuvette. Sheath fluid is expelled into a well
Prime
Primes the system to remove air from the
sheath side of the instrument.