Luminex 100 IS Version 2.2 User Manual
Page 60
Luminex
100
IS Developer Guide to
x
MAP Technology Version 2.2
x
MAP
Technology
54
PN 89-00002-00-032 Rev. A
Figure 38. Template Setup Wizard Dialog Box—Select Data
Collection Only
3. Select Data Collection Only for the Template Type.
4. Enter the Sample Vol (uL) for this template. This is the amount
of sample that will be tested. The volume in the wells must be at
least 25 uL greater than the sample volume.
5. Enter a Sample Timeout (sec) for this template. This is the
maximum time in seconds that the analyzer will spend acquiring
each sample.
6. Enter the Low Limit and High Limit for the Doublet
Discriminator Gate.
7. Click Next. The next Template Setup Wizard dialog box
appears. See Figure 39.
Select Data Collection Only