Qualitative assay using standards or controls – Luminex 100 IS Version 2.2 User Manual
Page 51

x
MAP
Technology
Create Templates
PN 89-00002-00-032 Rev. A
45
17. After all the information has been entered for your template,
check it carefully. When all the information is correct, click Save
to save the template, then click Close to exit.
Note:
You can return to and edit any unused templates using the
Edit Template feature.
Qualitative Assay Using
Standards or Controls
To create a template for qualitative tests.
Note:
You must define each Qualitative template using a single
product. The product must contain one standard. The standard is
assigned a “Qual” value for each test. Control products are
optional, but recommended for judging the acceptability of batch
results.
1. Click the Create Template button on the Developer Workbench
toolbar. The Template Setup Wizard dialog box appears. See
Figure 29.
Sanitize
Acquires sanitizing solution from the reservoir
to sanitize or eliminate air in the cuvette.
Skip
Program a well skip into template
Verify CON 1
Acquire CON 1 beads from a well to verify
that the system is working within the
calibrated settings.
Verify CON 2
Acquire CON 2 beads from a well to verify
that the system is working within the
calibrated settings.
Warmup
Checks for proper system warm up and if
needed begins the warmup process.
Wash from
Reservoir
Acquires wash solution from reservoir to wash
the system.
Wash from Well
Acquires wash solution from a well to wash
the system.
Table 9: Available Commands (Continued)
Command
Definition