Luminex 100 IS Version 2.2 User Manual
Page 43
x
MAP
Technology
Create Products
PN 89-00002-00-032 Rev. A
37
Apply Suffix—this button applies the number one to the first
test name. For each successive test name it applies the test name
and number at one increment higher than the previous test name.
For example: Suffix = Protein, then Test 1 is Protein 1, Test 2 is
Protein 2, and so on.
Apply Unit—this button applies the same unit value that you
entered on the first line to all other tests. For example, if you
enter the mg/uL value for the first Test Unit, then select the
Apply Unit button to make all test units mg/uL.
7. Enter the Number of Standards (assay) for this product and
press Enter key.
8. Enter the Reagent Name for each standard in this product. Each
standard reagent must test for every analyte (bead set) used in the
assay.
Note:
If the reagent names are similar, click the Apply Suffix
button to ensure each name is unique.
9. Enter the Number of Controls (assay) for this product and press
Enter key.
10. Enter the Reagent Name for each control in this product. Each
control reagent must test for every analyte (bead set) used in the
assay.
Note:
If the reagent names are similar, click the Apply Suffix
button to ensure each name is unique.
11. After entering all the information, check it carefully. When all
the information is correct, click Save to save the product.
12. Click Close.