INSTALLATION GUIDE

Error Codes

(revised 5-25-2009)

Code Number

Description

Correction

Warnings

0001

Serial number not present

Hardware serial number has become inoperative – system performance will not be

influenced.

0010

Signal Strength is below Signal Strength Cutoff entry

Low signal strength is typically caused by one of the following:

» Empty pipe

» Improper programming/incorrect values

» Improper transducer spacing

» Non-homogeneous pipe wall

0011

Measured Speed of Sound in the liquid is greater than ±10% different

than the value entered during meter setup

Verify that the correct liquid was selected in the BASIC menu.
Verify that pipe size parameters are correct

Class C Errors

1001

System tables have changed

Initiate a meter RESET by cycling power or by selecting SYSTEM RESET in the SEC

MENU.

1002

System configuration has changed

Initiate a meter RESET by cycling power or by selecting SYSTEM RESET in the SEC

MENU.

Class B Errors

3001

Invalid hardware configuration

Upload corrected file

3002

Invalid system configuration

Upload corrected file

3003

Invalid strategy file

Upload corrected file

3004

Invalid calibration data

Re-calibrate the system

3005

Invalid speed of sound calibration data

Upload new data

3006

Bad system tables

Upload new table data

Class A Errors

4001

Flash memory full

Return unit to factory for evaluation

Table A-5.1 - Error Codes