beautypg.com

Installation guide, Target dbg data screen - definitions, Saving meter configuration on a pc – Veris Industries FSRxxxx SERIES Install User Manual

Page 34: Printing a flow meter configuration report

background image

FSRxxxx SERIES

Z205739-0D

PAGE 34

©2013 Veris Industries USA 800.354.8556 or +1.503.598.4564 / [email protected]

05131

Alta Labs, Enercept, Enspector, Hawkeye, Trustat, Aerospond, Veris, and the Veris ‘V’ logo are trademarks or registered trademarks of Veris Industries, L.L.C. in the USA and/or other countries.

TM

INSTALLATION GUIDE

The second step (Page 2 of 3) in the calibration process is the selection of the
engineering units with which the calibration will be performed. Select the Flow Rate
Units and click the Next button at the bottom of the window.

Next>

Cancel

File Open...

File Save...

Calibration (Page 2 of 3) - General Setup

Gallons

Min

Flow Rate Units:

/

It is advisable to File Save the existing calibration before modifying it. If the Flow Rate Units selected on this page do not

match the Flow Rate Units utilized for the existing data points collected on Page 3 of 3, flow measurement errors can occur.

To view measurement units, go to Page 3 of 3 and press Edit. The Calibration Points Editor will show what units

were used during the existing calibration.

1) If no data exists in the editor, selection of Flow Rate Units will not influence measurements.

2) If new calibration points are to be entered on Page 3 of 3, it is advisable to remove the existing calibration

points using the Calibration Points Editor.

Figure 5.9 - Calibration Page 2 of 3

Page 3 of 3 as shown in Figure 5.10 allows multiple actual flow rates to be recorded.
To calibrate a point, establish a stable, known flow rate (verified by a real-time
primary flow instrument), enter the actual flow rate in the Figure 5.10 window, and
click the Set button. Repeat for as many points as desired.

Note: If only two points are to be used (zero and span), it is preferable to use the highest flow rate
anticipated in normal operation as the calibration point. If an erroneous data point is collected,
remove it pressing the Edit button, selecting the bad point, and then selecting Remove.

Next>

Cancel

File Open...

File Save...

Calibration (Page 2 of 3) - General Setup

1) Please establish a

reference flow rate.

1FPS / 0.3MPS Minimum.

2) Enter the reference flow rate

below. (Do not enter 0)

3) Wait for flow to stablize.

4 Press the Set button.

G

al/MIN

Delta Time

Flow:

Export...

Edit

Figure 5.10 - Calibration Page 3 of 3

Zero values are not valid for linearization entries. Flow meter zero is entered on Page
1 of 3. The following error message appeard if a zero calibration point is attempted.

Calibration points are too close. Calibration not usable.

USP

OK

!

Press the Finish button when all points have been entered.

Target Dbg Data Screen - Definitions

1. Calc Count - The number of flow calculations performed since the meter was last

power cycled.

2. Sample Count - The number of samples taken per second.

3. Raw Delta T (ηs) - The duration of one ultrasonic pulse crossing the pipe.

4. Course Delta T - The meter uses 2 wave forms: a coarse wave to find the best delay

and other timing measurements, and a fine wave for the flow measurement.

5. Gain - The amount of signal amplification applied to the reflected ultrasound pulse

to make it readable by the digital signal processor.

6. Gain Setting/Waveform Power - The first number is the gain setting on the digital

pot (automatically controlled by the AGC circuit). Valid numbers are from 1 to 100.
The second number is the power factor of the current waveform. For example “8”
indicates a 1/8 power wave form.

7. Tx Delay - The amount of time the transmitting transducer waits for the receiving

transducer to respond before the transmitter initiates another measurement
cycle.

8. Flow Filter - The current value of the adaptive filter.

9. SS (Min/Max) - The minimum and maximum signal strength levels encountered by

the meter beginning at the time the meter was power cycled.

10. Signal Strength State - Indicates whether the present signal strength minimum

and maximum are within a pre-programmed signal strength window.

11. Sound Speed - The actual sound speed measured by the transducers.

12. Reynolds - A number indicating how turbulent a fluid is. Reynolds numbers

between 0 and 2000 are considered laminar flow. Numbers between 2000 and
4000 are in transition between laminar and turbulent flows. Numbers greater
than 4000 indicate turbulent flow.

13. Reynolds Factor - The value applied to the flow calculation to correct for variations

in Reynolds numbers.

14. Serial Number - The serial number reported by firmware.

Saving Meter Configuration on a PC

Save the complete configuration of the flow meter from the Configuration screen.
Select the File Save button located in the lower left-hand corner of the screen and
name the file. Files are saved as a *.dcf extension. This file may be transferred to
other flow meters or may be recalled should the same pipe be surveyed again or
multiple meters programmed with the same information.

Printing a Flow Meter Configuration Report

Select File from the upper task bar and Print to calibration/configuration information
sheet for the installation.