beautypg.com

Installation guide, Ser menu -- service menu – Veris Industries FSRxxxx SERIES Install User Manual

Page 25

background image

FSRxxxx SERIES

Z205739-0D

PAGE 25

©2013 Veris Industries USA 800.354.8556 or +1.503.598.4564 / [email protected]

05131

Alta Labs, Enercept, Enspector, Hawkeye, Trustat, Aerospond, Veris, and the Veris ‘V’ logo are trademarks or registered trademarks of Veris Industries, L.L.C. in the USA and/or other countries.

TM

INSTALLATION GUIDE

SER Menu -- Service Menu

The SER MENU menu allows access to meter set up values that may need revision due
to application specific conditions and information valuable in troubleshooting.

SER MENU -- Service Menu:

1. SSPD MPS -- Liquid Sound Speed (Meters per Second) (Reported by Firmware)

2. SSPD FPS -- Liquid Sound Speed (Feet per Second) (Reported by Firmware)

The meter performs an actual speed of sound calculation for the liquid it is
measuring. This speed of sound calculation varies with temperature, pressure and
fluid composition.

The meter compensates for fluid sound speeds that vary within a window of ± 10%
of the liquid specified in the BSC MENU. If this range is exceeded, error code 0011
appears on the display and the sound speed entry must be corrected.

The value indicated in SSPD measurement should be within 10% of the value
entered/indicated in the BSC MENU item FLUID SS. (The SSPD value itself cannot be
edited.) If the actual measured value is significantly different (> ± 10%) than the BSC
MENUs FLUID SS value, it typically indicates a problem with the instrument setup. An
entry such as FL TYPE, PIPE OD or PIPE WT may be in error, the pipe may not be round,
or the transducer spacing is not correct.

Table 4.5 lists sound speed values for water at varying temperatures. If the meter is
measuring sound speed within 2% of the table values, then the installation and setup
of the instrument is correct.

Temp.

Velocity

Temp.

Velocity

Temp.

Velocity

°C

°F

MPS

FPS

°C

°F

MPS

FPS

°C

°F

MPS

FPS

0

32

1402

4600

80

176

1554

5098

160

320

1440

4724

10

50

1447

4747

90

194

1550

5085

170

338

1412

4633

20

68

1482

4862

100

212

1543

5062

180

356

1390

4560

30

86

1509

4951

110

230

1532

5026

190

374

1360

4462

40

104

1529

5016

120

248

1519

4984

200

392

1333

4373

50

122

1543

5062

130

266

1503

4931

220

428

1268

4160

60

140

1551

5089

140

284

1485

4872

240

464

1192

3911

70

158

1555

5102

150

302

1466

4810

260

500

1110

3642

Table 4.5 - Sound Speed of Water

3. SIG STR -- Signal Strength (Reported by Firmware)

The SIG STR value is a relative indication of the amount of ultrasound making it
from the transmitting transducer to the receiving transducer. The measurement of
signal strength assists service personnel in troubleshooting the system. In general,
expect the signal strength readings greater than 5 on a full pipe with the transducers
properly mounted. Signal strength readings that are less than 5 indicate a need to
choose an alternative mounting method for the transducers or that an improper pipe
size has been entered.

Signal strength below the Low Signal Cutoff (SIG C-OF) value will generate a 0010
error (Low Signal Strength) and require either a change in the SIG C-OF value or
transducer mounting changes.

Note: If the unit is configured to display totalizer values, the display will alternate between ERROR
0010 and the totalizer value.

Signal strength readings in excess of 98 indicate that a longer path length between
the transducers is needed. For example, if transducers mounted on a 3 inch PVC
pipe in V-Mount cause the measured signal strength value to exceed 98, change the
mounting method to W-Mount for greater stability in readings.

Because signal strength is not an “absolute” indication of how well a meter is
functioning, there is no real advantage to a signal strength of 50 over a signal
strength of 10.

TEMP 1 -- Temperature of RTD 1 (Reported by Firmware in °C): When RTD is
selected from the CH2 menu and RTDs are connected to the BTU meter, the
firmware displays the temperature measured by RTD 1 in °C.

TEMP 2 -- Temperature of RTD 2 (Reported by Firmware in °C): When RTD is
selected from the CH2 menu and RTDs are connected to the BTU meter, the
firmware displays the temperature measured by RTD 2 in °C.

TEMPDIFF -- Temperature difference (Reported by Firmware in °C): When RTD
is selected from the CH2 menu and RTDs are connected to the BTU meter, the
firmware displays the difference in temperature measured between RTD 1
and RTD 2 in °C.

4. SIG C-OF -- Low Signal Cutoff (Value): 0.0 - 100.0

The SIG C-OF is used to drive the flow meter and its outputs to the SUB FLOW
(Substitute Flow described below) state if conditions occur that cause low signal
strength. A signal strength below 5 is generally inadequate for measuring flow
reliably, so the minimum setting for SIG C-OF is 5. A good practice is to set the SIG
C-OF at approximately 60-70% of actual measured maximum signal strength.

Note: The factory default “Signal Strength Cutoff” is 5.

If the measured signal strength is lower than the SIG C-OF setting, error 0010 appears
on the display until the measured signal strength increases to greater than the cutoff
value.

A signal strength indication below 2 is considered to be no signal at all. Verify that
the pipe is full of liquid, the pipe size and liquid parameters are entered correctly and
that the transducers are mounted accurately. Highly aerated liquids cause low signal
strength conditions.