SITRANS P measuring instruments for pressure

Remote seals

Measuring setups with remote seals

2/184

Siemens FI 01 · 2009

2

Types of installation for level measurements (closed vessels)

H

O

H

V

Installation type E
Start-of-scale: p

MA

= r

FL

· g · H

U

- r

Oil

· g · H

V

Full-scale:

p

ME

= r

FL

· g · H

O

- r

Oil

· g · H

V

Legend
p

MA

Start-of-scale value to be set

p

ME

Full-scale value to be set

r

FL

Density of medium in vessel

r

Oil

Density of filling oil in the capillary to

the remote seal

g

Local acceleration due to gravity

H

U

Start-of-scale value

H

O

Full-scale value

H

V

Distance between the measuring points (spigots)

Installation type E

DH

H

U

H

U

DH

H

O

H

V

H

2

H

1

Installation type G, H and J
Start-of-scale: p

MA

= r

FL

· g · H

U

- r

Oil

· g · H

V

Full-scale:

p

ME

= r

FL

· g · H

O

- r

Oil

· g · H

V

Legend
p

MA

Start-of-scale value to be set

p

ME

Full-scale value to be set

r

FL

Density of medium in vessel

r

Oil

Density of filling oil in the capillary to

the remote seal

g

Local acceleration due to gravity

H

U

Start-of-scale value

H

O

Full-scale value

H

V

Distance between the measuring points (spigots)

H

1

H

2

Installation type J

between the measuring points, no vacuum

H

2

£ 7 m (23 ft), with halocarbon oil as

filling liquid only H

2

£ 4 m (13.1 ft)

H

1

H

2

Installation type H

below the lower measuring point

Installation type G

Pressure transmitter for differential pressure

above the upper measuring point, no vacuum

H

1

£ 7 m (23 ft), with halocarbon oil as

filling liquid only H

1

£ 4 m (13.1 ft)

© Siemens AG 2008