SERVICE MANUAL

TNC

Issue:

01.09.1995

Page 75

When the override potentiometer is turned to the right, the control outputs an analog voltage which is increased

proportionally to the servo lag up to

max. The control is operating correctly if a voltage of

can be measured at the test adaptor with the multimeter. If no voltage can be measured, switch

off the main switch, disconnect the nominal value line from the servo amplifier and test for short-circuit. If the

nominal value line is in order, connect the connector X8 to the LE again (leave the nominal value line dis-

connected), switch on the main switch and repeat the measurement with reference mark traverse. If an
analog voltage can be measured now, the control is operating correctly. If no voltage can be measured,
the analog output of the LE is probably defective.

15.2.1 Checking

the Analog Outputs:

Setup

Test Adaptor

Multimeter

X8

Nominal

Value Output

Flange socket with female insert