Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual
Page 98

94
Reference
Welch Allyn Propaq CS Vital Signs Monitor
HR Limits
Adult: 50, 120 beats per minute
Ped: 50, 150 beats per minute
Neonate: 100, 200 beats per minute
NIBP Limits - Systolic
Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg
NIBP Limits - Diastolic
Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
NIBP Limits - Mean
Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
P1, P2 Limits - Systolic
Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg
P1, P2 Limits - Diastolic
Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
P1, P2 Limits - Mean
Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
SpO
2
Limits
Adult: 90%, 100%
Ped: 90%, 100%
Neonate: 85%, 95%
RR/BR
Adult: 5, 30 Br/M
Ped: 10, 45 Br/M
Neonate: 10, 75 Br/M
TEMP Limits
35.0°, 37.8° C
ΔT Limits
0.0°, 2.8° C
ETCO
2
Limits
25, 60 mmHg (3.0 and 8.0 for % and kPa)
INCO
2
Limits
N/A, 5 mmHg (0.7 for % and kPa)
Apnea Delay
Adult/Ped: 20 seconds
Neonate: 15 seconds
Printer Settings
Printer Alarm Print
OFF
Printer Auto Print
OFF
Printer NIBP Ticket
OFF
Printer Apnea Ticket
ON
Printer Print Speed
25 mm/s
Printer Auto Trend
OFF
Printer Trend Selections
NIBP and P1 = ON; all others = OFF
Printer OxyCRG on Alarm
OFF
a.
Any time you change the Date, Filter, Temp F/C, Decimal, HR/PR Alarm Limits (Can or Cannot Turn Off) or CO
2
Units setting, the new
setting also becomes the powerup default setting.
Setting
Factory default