beautypg.com

Indications during a 90-second silence period – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual

Page 54

background image

50

Alarms & Alerts

Welch Allyn Propaq CS Vital Signs Monitor

Re-enable an alarm or alert tone before the 90-second silence period has
elapsed

1.

Press Silence/Reset.

Indications during a 90-second silence period

At the Propaq CS monitor

At the Acuity Central Station

During the silence period, the usual visual alarm and alert indications are displayed.

If a new alarm or alert occurs, the new alarm or alert
tone interrupts the silence period.

Only life-threatening arrhythmia and apnea alarm tones
interrupt the suspend period.

To re-enable an alarm or alert tone before the
90-seconds has elapsed, press Silence/Reset.

To resume an alarm or alert tone before the 90-seconds has
elapsed, press RESUME.

If an equipment alert occurs during the silence period, you can acknowledge (dismiss) all indications of the alert by
pressing any key, as usual.

If the original alarm or alert condition still exists after 90 seconds, the alarm or alert tone sounds again.