Monitor (physical) specifications – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual
Page 114

110
Reference
Welch Allyn Propaq CS Vital Signs Monitor
Monitor (physical) specifications
Characteristic
Specification
Protection classifications, all configurations
a
a.
Per EN 60601-1 unless otherwise stated.
Type of Protection against Electric
Shock—Power Adapter
Power adapter class 1
Type of Protection against Electric
Shock—Monitor (connected to power
adapter or internal battery)
Protective earth not available in monitor. Monitor designed and tested to meet
Double Insulation Requirement.
Degree of Protection Against Electric
Shock, for Parts Applied to Patients
See monitor labels
Recovery time following defibrillator
discharge
Less than or equal to 10 seconds
Electrosurgery interference
suppression
Suitable for use in the presence of electrosurgery
Method of Disinfection
Not suitable for autoclaving (see cleaning instructions,
Flammable Anesthetics
Not suitable for use with flammable anesthetics
Monitor only
Height
8.2 in (20.8 cm) with handle
Width
9.6 in (24.4 cm)
Depth
5.6 in (14.1 cm)
Weight
7.6 lb (3.4 kg)
Monitor with SpO
2
module
Height
8.2 in (20.8 cm) with handle
Width
9.6 in (24.4 cm)
Depth
7.7 in (19.7 cm)
Weight
10.8 lb (4.9 kg)
Monitor with expansion module
(Printer / SpO
2
/ MCO
2
)
Height
11.4 in (28.8 cm)
Width
9.6 in (24.4 cm)
Depth
7.7 in (19.7 cm) with back feet
Weight with Printer, SpO
2
, and MCO
2