beautypg.com

Power adapter, Specifications, Power adapter specifications – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual

Page 117

background image

Directions for Use

Reference

113

Power Adapter Specifications

Characteristic

Specification

Protection classifications, all adapters

a

b

a.

Per EN 60601-1 unless otherwise stated.

b.

See Welch Allyn Products and Accessories for model numbers.

Type of Protection Against Electric Shock

Class I, (Protectively Earthed)

Degree of Protection Against Harmful Ingress
of Water

For ordinary, indoor locations only.

Method of Disinfection

Not suitable for autoclaving

Flammable Anesthetics

Not suitable for use with flammable anesthetics

Environmental specifications, all adapters

Operating Temperature

0° to 50° C

Shipping and Storage Temperature

-20° to 60° C

Operating Altitude

-2,000 to 15,000 feet (-610 to 4,572 m)

Shipping and Storage Altitude

-2,000 to 40,000 feet (-610 to 12,192 m)

Operating Relative Humidity

15% to 95%, noncondensing

Shipping, Storage Relative Humidity

15% to 95%, noncondensing

Shock

50 g

Vibration

Random Vibration, 0.02 g

2

/Hz from 10 to 300 Hz, ramping down to 0.002

g

2

/Hz at 500 Hz. Operating 1 hour per axis, 3 hours/ test.

Physical specifications

Length

5.0 in (12.7 cm)

Width

3.6 in (9.1 cm)

Height

3.1 in (7.9 cm)

Weight

3.1 lb (1.4 kg)

Universal power adapter, part No. 503-0054-00

Rated Input

100-120 V ac, 500 mA, 50/60 Hz

Rated Fuses

T800 mA/250 V, Time-Delay, 5x20mm

Rated Output (Continuous)

16-24 V dc, 25 VA

Additional Features

Detachable power cord, pilot light

Universal power adapter, part No. 503-0054-01

Rated Input

200-240 V ac, 250 mA, 50/60 Hz

Rated Fuses

T400 mA/250 V, Time-Delay, 5 x 20mm

Rated Output (Continuous)

16-24 V dc, 25 VA

Additional Features