Setup

Prepare the Propaq CS monitor for a new patient

1.

Press the gray recessed MONITOR button on the right side of the monitor to turn the
monitor off (if it is on). Press it again to turn the monitor on. The monitor displays the
powerup screen for about 10 seconds, then displays the Main Menu. The monitor is in
the powerup patient mode with the associated settings.

2.

Confirm that the monitor emits a tone. If the monitor has SpO

2

, listen for two tones

and confirm both speakers are working.

3.

Confirm the monitor is in the correct patient mode according to the patient’s age. If
the patient mode is not correct, from the Main Menu press SETUP, MORE, CHANGE
to access the Patient Mode window:

4.

Based on the patient’s age, press NEONATAL, PEDIATRIC, or ADULT. When the
confirmation window appears, press YES to confirm your selection.

WARNING Before you use a Propaq CS monitor on a new patient, always turn it
off for a few seconds and then turn it on again. This clears the prior patient’s trend
values, alarm limit settings, and NIBP cuff inflation target.

Note

Check the battery voltage level on the powerup screen (or check it on
the Time/Day window: Home, SETUP, MORE, MORE). If the battery
voltage is 7.4V or less or a low battery message is displayed, connect
the monitor to an ac power adapter to recharge the battery (page 20).
Connecting the adapter does not interrupt patient monitoring.

Home

PATIENT MODE

NEONATAL

PEDIATRIC

ADULT

SETUP

SELECT PATIENT MODE BASED ON AGE:

PED : > 44 WEEKS GEST. AGE,

ADULT:

NEO : < 44 WEEKS GEST. AGE

< 9 YEARS

> 9 YEARS

BATTERY: 8.2 VOLTS

PREVIOUS

MENU