beautypg.com

Reference, Set the time and date, 8 - reference – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual

Page 95: Time/day settings and trends

background image

8

91

Reference

Set the time and date

1.

From the Main Menu press SETUP, MORE, MORE to display the Time/Day window:

2.

Press NEXT, UP, and DOWN as needed to set the time and date. Then press ENTER
to store the new time and date.

Time/Day settings and trends

When you change the hour/minute/second setting for the monitor in the Time/Day
window, the monitor deletes any patient trend data that is older than five hours for non-
NIBP trends or older than eight hours for NIBP trends according to the new clock setting.

However, if the monitor has not yet stored the full capacity of trends and you change the
hour/minute/second setting to a time that is within the stored trend period, previously
stored trends are not erased.

Changing the day, month, or year setting does not affect the stored patient trends.

BR

TEMP

100.4

98.6

1.8

97

( 85)

mmHg

(M)

%

F

58

MCO

2

S

p

O

2

mmHg

38

Br/m

D

NEXT

T1
T2
ΔT

SERVICE

TIME

DAY

UP

DOWN

ENTER

H:MIN:S

15

:55:10

MO/DA/YR

10/21/99

12

WARNING Changing the hour/minute/second setting for the monitor in the
Time/Day window can cause the monitor to erase previously stored patient trend
data.