Change the powerup patient mode – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual
Page 23

Directions for Use
Setup
19
Change Powerup Patient Mode or store customized
settings
The Propaq CS monitor has standard, preset, default powerup settings and alarm limits for
each patient mode: Adult, Pediatric, and Neonatal. These are “Factory Patient Mode”
settings (listed on
“Factory default settings” on page 93
You can also choose to customize and store programmable powerup settings and alarm
limits for each patient mode. These are “Custom Patient Mode” settings.
The instructions below describe how to change the powerup patient mode and how to
select and store new Custom Patient Mode settings.
Change the Powerup Patient Mode
1.
From the Main Menu, press SETUP, MORE, CHANGE, SETUP to access the Mode
Setup window.
2.
Press NEXT to highlight the desired Factory or Custom powerup mode, then press
POWERUP* and YES. The new powerup selection is marked by the asterisk.
Changing the powerup mode does not affect the patient mode currently used.
Note
When you change patient modes, you also change the alarm limits associated
with the new patient mode.
MODE SETUP
BR
TEMP
100.4
98.6
1.8
97
P
2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
S
D
NEXT
T1
T2
ΔT
PREVIOUS
MENU
POWERUP*
USE
NOW
FACTORY:
SAVE
BATTERY: 9.2 VOLTS
*ADULT
PED
CUSTOM
* = PATIENT MODE ON POWERUP
ADULT
PED
NEO
NEO
:
12
Asterisk indicates which patient
mode is currently selected for
powerup.