Monitoring
25
13. Press MORE to display the second ECG/RESP menu and status window:
0
14. Press NEXT and CHANGE as desired to adjust the display.
ECG1 SIZE
Selects the ECG1 waveform size: 4, 2, 1, 0.5, or 0.2 mV/cm.
ECG1 LEAD Selects the ECG1 lead: I, II, III, aVR, aVL, aVF, or V.
Selections aVR, aVL, aVF, and V are only available with a 5-lead ECG
cable.
The ECG2 lead is always V, except when the ECG1 lead is V (in that
case the ECG2 lead is II).
ECG2 SIZE
Selects the ECG2 waveform size: 4, 2, 1, 0.5, or 0.2 mV/cm.
RESP SIZE
Selects the RESP waveform size: 1x, 2x, 4x, 8x, or 16x.
The QRS detector sensitivity threshold is not affected by changing the
ECG display size. Likewise, the RESP breath detector threshold is not
affected by changing the RESP display size.
HR/PR TONE
Sets heart tone loudness to LOW, MEDIUM, HIGH, or OFF.
If SpO
2
is monitored, tone pitch varies with the SpO
2
value.
PACER DISPLAY
Turns on and off the pacer indicator in the ECG waveform.
If the patient has a pacemaker, you may want to turn on
the pacer indicator (see page 27).
ECG BANDWIDTH
Selects the bandwidth for displayed and printed data.
MONITOR is 0.5 Hz to 40 Hz (Adult mode) or 0.5 Hz to 120 Hz
(Pediatric and Neonatal mode).
Monitor Mode filters out extraneous noise and artifact
to provide a more stable display.
EXTENDED is 0.05 Hz to 40 Hz (Adult mode) or 0.05 Hz to 120
Hz (Pediatric and Neonatal mode).
Extended Mode is a higher-resolution setting that allow
more detailed analysis.
12
ECG/RESP
MON
BR
( 96)
mmHg
(M)
TEMP
100.4
98.6
1.8
97
121
79
S
D
P
2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
S
D
NEXT
CHANGE
PREVIOUS
MENU
T1
T2
ΔT
mV
cm
1
: ON
: MONITOR (0.5-40Hz)
: Ld2 (RA-LL)
:
OFF
HR/PR TONE
PACER DISPLAY
ECG BANDWIDTH
RESP LEAD
RESP MONITORING : ON
ECG bandwidth selected:
MON = Monitor
EXT = Extended