Page 69: Print the displayed waveforms

background image

5

65

Printing & trends

Print patient data

Print the displayed waveforms

Press SNAPSHOT or START/STOP. The Propaq CS monitor prints up to three of the
displayed waveforms.

When four waveforms are displayed, the monitor prints the top three displayed
waveforms (except for the ECG2 waveform which is never printed).

If you pressed START/STOP, the monitor continues to print until you press START/
STOP
again.

Note

START
STOP

SNAP
SHOT

PRINT
TRENDS

Manually starts and stops a printout of patient
information as it is monitored (continuous or real time).

Prints all trends that are enabled in the Printer Setup
Page (see page 67).

Prints the last eight seconds of data for nonrespiration
waveforms and 32 seconds of compressed waveform
history for respiration waveforms.

The numerics across the top of the printout apply to the
moment you press SNAPSHOT.

Printer Buttons

The symbol indicates that the NIBP reading was taken in the presence of high
artifact while monitoring ECG with the Smartcuf motion artifact filter on. Artifact
can affect accuracy. To help reduce artifact, see