20

Setup

Welch Allyn Propaq CS Vital Signs Monitor

Customize Patient Mode settings

1.

From the Main Menu, press SETUP, MORE, CHANGE, SETUP to access the Mode
Setup window.

2.

The patient mode you want to reprogram (ADULT, PED, or NEO) must be currently
selected. To make sure it is currently selected, press NEXT as needed to highlight the
desired mode, then press USE NOW and YES.

3.

Press Home to exit the Mode Setup window, then use other menus and keys
to set the monitor settings and alarm limits as desired.

A convenient way to access settings and alarm limits for all functions without
connecting cables is to select the in-service mode (disconnect all patient cables, turn
the monitor power off and then on, then press SETUP, WAVE SELECT, INSERVICE
from the Main Menu).

4.

Re-enter the Mode Setup window, press NEXT as needed to highlight the desired
Custom mode, then press SAVE and YES.

If the in-service mode is used, turn off the monitor to exit the in-service mode.

WARNING If any alarms are set to OFF and you select SAVE to store settings for
a Custom patient mode, those alarms will be OFF when the monitor powers up in
that Custom patient mode or that Custom patient mode is selected. Consider
carefully before setting Custom patient mode powerup alarms to OFF.