Change the date format, ecg filter, and units, Change the date format, ecg filter, and, Units – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual
Page 96

92
Reference
Welch Allyn Propaq CS Vital Signs Monitor
Change the date format, ECG filter, and units
1.
Make sure you are in the Adult patient mode (from the Main Menu press SETUP,
MORE, CHANGE, ADULT, YES).
2.
From the Main Menu press SETUP, MORE, MORE, SERVICE, YES (to access the
Service Menu), MORE, MORE, SETTINGS. The monitor displays the Settings
window:
3.
Press NEXT and CHANGE to select the desired settings.
DATE
Sets the date format: Month/Day/Year, Day.Month.Year, or Year/Month/Day.
FILTER
Sets the ECG filter frequency: 60 Hz, 50 Hz, or OFF. Make sure it is set to
your ac mains frequency.
TEMP F/C
Sets the temperature display units: Fahrenheit or Celsius. Changing units
does not erase the TEMP trends.
DECIMAL
Sets the decimal character as either a period (.) or a comma (,).
HR/PR
ALARM
LIMITS
Allows or prohibits turning off the HR/PR alarm limits. If CANNOT TURN
OFF is selected, the ON/OFF key is not displayed for HR/PR in the Alarm
Limits Menu.
CO
2
UNITS
Sets the CO
2
display units as mmHg, kPa, or percent (%).
Changing units erases the CO
2
trends and changes CO
2
alarm limit
settings to the factory default settings for the currently-used patient mode.
Note
Any time you change the Date, Filter, Temp F/C, Decimal, HR/PR Alarm Limits
(CAN or CANNOT TURN OFF), or CO
2
Units setting, the new setting also
becomes the powerup default setting.
SETTINGS
BR
TEMP
100.4
98.6
1.8
97
P
2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
S
D
NEXT
T1
T2
ΔT
PREVIOUS
MENU
BATTERY: 9.3 VOLTS
DATE
CHANGE
FILTER
TEMP F/C :
DECIMAL : .
HR/PR ALARM LIMITS: CAN TURN OFF
:
:
MO/DA/YR
60 Hz
C
CO2 UNITS: mmHg
12