Change the date format, ECG filter, and units

1.

Make sure you are in the Adult patient mode (from the Main Menu press SETUP,
MORE, CHANGE, ADULT, YES).

2.

From the Main Menu press SETUP, MORE, MORE, SERVICE, YES (to access the
Service Menu), MORE, MORE, SETTINGS. The monitor displays the Settings
window:

3.

Press NEXT and CHANGE to select the desired settings.

DATE

Sets the date format: Month/Day/Year, Day.Month.Year, or Year/Month/Day.

FILTER

Sets the ECG filter frequency: 60 Hz, 50 Hz, or OFF. Make sure it is set to
your ac mains frequency.

TEMP F/C

Sets the temperature display units: Fahrenheit or Celsius. Changing units
does not erase the TEMP trends.

DECIMAL

Sets the decimal character as either a period (.) or a comma (,).

HR/PR
ALARM
LIMITS

Allows or prohibits turning off the HR/PR alarm limits. If CANNOT TURN
OFF is selected, the ON/OFF key is not displayed for HR/PR in the Alarm
Limits Menu.

CO

2

UNITS

Sets the CO

2

display units as mmHg, kPa, or percent (%).

Changing units erases the CO

2

trends and changes CO

2

alarm limit

settings to the factory default settings for the currently-used patient mode.

Note

Any time you change the Date, Filter, Temp F/C, Decimal, HR/PR Alarm Limits
(CAN or CANNOT TURN OFF), or CO

2

Units setting, the new setting also

becomes the powerup default setting.

SETTINGS

BR

TEMP

100.4

98.6

1.8

97

P

2

( 15)

mmHg

(M)

25

9

S
D

NIBP

( 85)

mmHg

(M)

%

F

122

58

MCO

2

S

p

O

2

mmHg

38

Br/m

S
D

NEXT

T1
T2
ΔT

PREVIOUS

MENU

BATTERY: 9.3 VOLTS

DATE

CHANGE

FILTER

TEMP F/C :

DECIMAL : .

HR/PR ALARM LIMITS: CAN TURN OFF

:

:

MO/DA/YR

60 Hz

C

CO2 UNITS: mmHg

12