Set printer options and automatic printing – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual
Page 72

68
Printing & trends
Welch Allyn Propaq CS Vital Signs Monitor
Set printer options and automatic printing
1.
Press SETUP, MORE, PRINTER from the Main Menu to display the Printer Setup
Page:
2.
Press NEXT and CHANGE as desired to set printer options.
CONTINUOUS
Set the speed for continuous printing: 6.25, 12.5, or 25.0 mm/s.
AUTO PRINT
Automatically print a waveform snapshot at the specified interval:
15 or 30 minutes, or 1, 2, or 4 hours (or OFF).
ALARM PRINT
If ON, automatically prints patient data whenever a patient alarm
occurs, beginning with 12 seconds of patient data history stored
before the alarm occurred.
Printing continues for 20 seconds after you suspend the alarm. To
immediately stop printing, press START/STOP.
Note
Because the Alarm Print begins with the 12 seconds of patient data stored before
the alarm occurred, the monitor stores and prints all Alarm Print data 12 seconds
after the patient data appears on the display. The time annotated on the Alarm
Print indicates the time the data was recorded.
NIBP
TICKET
If ON, automatically prints an NIBP TICKET with NIBP data whenever
NIBP is measured.
APNEA
TICKET
If ON, automatically prints an APNEA TICKET with apnea data after the
patient resumes breathing and/or every minute the apnea alarm
continues.
OXYCRG
ON
ALARM
If ON, automatically prints an OxyCRG whenever an SpO
2
, HR/PR, RR/
BR, or apnea patient alarm occurs (see page 69).
PRINTER
BR
TEMP
100.4
98.6
1.8
97
P
2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
S
D
NEXT
T1
T2
ΔT
PREVIOUS
MENU
SETUP PAGE
AUTO TREND : 01 05 09 13 17 21
CONTINUOUS
AUTO PRINT
ALARM PRINT
NIBP TICKET
APNEA TICKET
OXYCRG ON ALARM
NIBP :
ON
RESP : OFF
P1
: ON
: 25.0 mm/s
: OFF
: OFF
: OFF
: ON
: OFF
P
2
: OFF
CHANGE
TREND
TEMP : OFF
12
Press to print all
selected trends.
Press to scroll to the
next selection.
Press to change the
displayed value.