beautypg.com

Indefinitely suspend all alarm and alert tones, Resume all alarm and alert tone capability, Indications during an all alarms suspended period – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual

Page 56

background image

52

Alarms & Alerts

Welch Allyn Propaq CS Vital Signs Monitor

Inhibit alarm and alert tones indefinitely: ALL ALARMS

You can suspend potential or current alarm and alert tones for an indefinite period of time.
The tones are disabled until someone re-enables them.

Indefinitely suspend all alarm and alert tones

1.

Press Home, SETUP, MORE, MORE, SERVICE, YES, MORE, MORE, SETTINGS,
ALL ALARMS, YES.

Resume all alarm and alert tone capability

1.

Press Silence/Reset.

Indications during an all alarms suspended period

Note

This feature requires access to the Service Menu. The Service Menu features
are not intended for use during ordinary, routine operation.

WARNING Whenever audible alarm tones are disabled, make sure the patient
is closely observed.

At the Propaq CS Monitor

At the Acuity Central Station

If an alarm or alert occurs during the suspend period, the usual visual alarm and alert indications are displayed.

Propaq monitors connected to Acuity Central Station:
Apnea or life-threatening arrhythmia alarm tones
interrupt the suspend period.

Standalone Propaq monitors: Apnea alarms do not
interrupt the suspend period, and arrhythmias are not
detected.

Apnea or life-threatening arrhythmia alarm tones interrupt the
suspend period.

To resume alarm and alert tone capability, press
Silence/Reset.

To resume alarm and alert tone capability, press RESUME.

If an equipment alert occurs during the suspend period, you can acknowledge (dismiss) all indications of the alert by
pressing any key, as usual.