Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual
Page 20

16
Setup
Welch Allyn Propaq CS Vital Signs Monitor
Whenever you change the patient mode, the alarm limit settings, maximum NIBP cuff
inflation pressures, and internal computations are automatically changed to the
defaults for that patient mode. See page 19 for information about preset Factory
patient modes or programmable Custom patient modes.
5.
To select which vital sign waveforms will display, from the Main Menu press SETUP,
MORE, WAVE SELECT. Use NEXT and ON/OFF to turn on the desired waveforms in
the Wave Select window.
You can turn on all waveforms, but only the first four waveforms selected as ON in the
Wave Select window are displayed. You cannot turn off the ECG1 waveform.
6.
To set the HR/PR source, display sweep speed, tone volumes, and display brightness,
from the Main Menu press SETUP, MORE to access Setup Menu 2. Use NEXT and
CHANGE to select settings.
Note
If you change the patient mode, the CO
2
alarm limits in the new mode might vary
slightly from the originally-programmed CO
2
alarm limits for the new mode.
Check the CO
2
alarm limits.
CURRENT SOURCE When the selected HR/PR source is no longer available, the current
source is the active source with highest priority. The RR/BR source
cannot be manually selected. It is always CO
2
if CO
2
is active.
Otherwise, it is ECG/RESP.
SELECTED SOURCE The user-selected HR/PR source is displayed along with the HR/PR
source currently used by the monitor.
SETUP
ECG
1
: ON
WAVE SELECT
CO
2
:
ON
RESP : ON
SpO
2
: ON
NIBP : ON
ECG
2
:
ON
NEXT
ON/
OFF
INSERVICE
PREVIOUS
MENU
PA
: ON
P
2
: ON
SETUP
CURRENT SOURCE : ???
SELECTED SOURCE : ECG
SWEEP (mm/s)
ALARM TONE
HR/PR TONE
PATIENT MODE
BRIGHTNESS
HR/PR RR/BR
NEXT
CHANGE
WAVE
SELECT
MORE
???
---
:
25.0
6.25
:
:
:
:
LOW
OFF
ADULT
NORMAL
PRINTER
? ? ? is displayed
until a source is
active.