beautypg.com

Temperature messages, Messages, Mainstream co – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual

Page 63

background image

Directions for Use

Alarms & Alerts

59

Temperature messages

PROBE NOT DETECTED. This message occurs when the Propaq CS monitor has
successfully measured temperature and a probe is then disconnected. Reconnect the
probe or acknowledge the equipment alert by pressing any menu key.

PROBE SHORT. Verify that the probe is properly inserted in the left side panel. If so,
replace probe.

CALIBRATION ERROR, TEMP DISABLED. This message appears when the Propaq
CS monitor has detected that it cannot accurately measure the temperature. The
monitor should be serviced.

Malfunction of the temperature probes may result in inaccurate readings. Confirm
suspect readings.

SpO

2

messages

SpO

2

messages can appear in the equipment alert window or in the SpO

2

numeric

window.

DEFECTIVE SPO2 SENSOR. If you see the error message DEFECTIVE SpO2
SENSOR, either the sensor is not compatible with the monitor or the sensor is not
working properly. Check Welch Allyn Products and Accessories (810-0409-XX) to be
sure the sensor is compatible. If compatibility is not a problem, try another sensor.

NO SENSOR DETECTED. Indicates an SpO

2

sensor has been disconnected from the

monitor after being plugged in for more than a few seconds.

SEARCH: During this search time, the SpO

2

channel tries to detect blood pulsing

through the measurement site. After the measurement has been established, the
oxygen saturation value is displayed in the numeric window.

STANDBY is displayed in the numeric window when the SpO

2

sensor is disconnected

from the patient, an alarm occurs, and you press the STANDBY key. STANDBY is also
displayed if you first plug the SpO

2

sensor cable into the monitor connector before

attaching the SpO

2

sensor to the patient.

Mainstream CO

2

messages

Messages for the mainstream CO

2

option can appear in the equipment alert window and

in numeric zones. If a sensor is damaged, contact Welch Allyn Technical Services
Department for information on sensor service options.

ALTIMETER FAILURE - RANGE. The Propaq CS monitor is operating at an altitude
outside the mainstream CO

2

option’s operating altitude range of -2,000 to 15,000

feet. Returning the monitor to within this range automatically cancels this message
and restores operation.

ALTIMETER FAILURE - RATE. The altimeter has detected that the ambient pressure is
changing at a rate greater than 100 mmHg/minute. When the rate of change is back
within the 100 mmHg/minute range, disconnect and reconnect the CO

2

sensor to the

Propaq CS monitor.

DEGRADED WAVEFORM - CHECK ADAPTER (UNCAL appears in the numerics area).
The mainstream CO

2

adapter is obstructed or the CO

2

sensor has failed. The CO

2