beautypg.com

Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual

Page 28

background image

24

Monitoring

Welch Allyn Propaq CS Vital Signs Monitor

7.

Plug the ECG cable into the ECG connector on the monitor's left side panel.

8.

Support the ECG cable so it does not stress the electrode wires, ECG cable
connectors, or electrodes.

9.

If an electrosurgical unit will be used, place the ECG cable and electrode wires as far
as possible from the surgical site and from the electrosurgical return electrode and its
cables. This minimizes interference.

Although the ECG channel contains electrosurgical interference suppression (ESIS)
circuitry, noise artifact may be displayed on the ECG trace while an electrosurgical
device is in use. Choose electrode placement to minimize interference.

10. Look for an ECG waveform and heart rate on the monitor. Depending on how your

monitor is programmed, a beep tone may occur with each detected QRS.

If there is no waveform, check the electrodes, wires, cable, and the monitor for a
possible misconnection or lead fault.

11. To set up the ECG/RESP display, from the Main Menu press ECG/RESP to display the

first ECG/RESP menu:

12. Press buttons as desired to adjust the display:

One of six V lead
electrode placement
sites for the C lead.

LL

LA

RA

RL

5-lead Configuration

V6

V1

Neonate 3-lead Configuration

RA

LA

LL

NET

OFF

HR

MON

MON

17:05:10

PRINT FAULT

II

BR

P

1

( 96)

BPM

mmHg

(M)

TEMP

Ld

2

100.4

98.6

1.8

97

121

79

S
D

P

2

( 15)

mmHg

(M)

25

9

S
D

NIBP

( 85)

mmHg

(M)

%

F

122

58

MCO

2

S

p

O

2

mmHg

38

Br/m

12

S
D

ECG

1

SIZE

ECG

1

LEAD

ECG

2

SIZE

RESP

SIZE

MORE

ADULT
ON NETWORK

T1
T2
ΔT

mV

cm

1

112

RESP

2

X

mV

cm

1

V

ALLEN KIM

ECG1 lead selected

ECG1 size selected

RESP waveform is displayed (with RESP

option)

RESP lead selected

RESP size selected

BR indicates respiration rate is derived from active CO

2

channel.

RR indicates CO

2

is not monitored and respiration rate is derived from ECG/RESP.

ECG2 lead

The ECG2 lead is always V, except when

the ECG1 lead is V (in that case the

ECG2 lead is II).