Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual
Page 28

24
Monitoring
Welch Allyn Propaq CS Vital Signs Monitor
7.
Plug the ECG cable into the ECG connector on the monitor's left side panel.
8.
Support the ECG cable so it does not stress the electrode wires, ECG cable
connectors, or electrodes.
9.
If an electrosurgical unit will be used, place the ECG cable and electrode wires as far
as possible from the surgical site and from the electrosurgical return electrode and its
cables. This minimizes interference.
Although the ECG channel contains electrosurgical interference suppression (ESIS)
circuitry, noise artifact may be displayed on the ECG trace while an electrosurgical
device is in use. Choose electrode placement to minimize interference.
10. Look for an ECG waveform and heart rate on the monitor. Depending on how your
monitor is programmed, a beep tone may occur with each detected QRS.
If there is no waveform, check the electrodes, wires, cable, and the monitor for a
possible misconnection or lead fault.
11. To set up the ECG/RESP display, from the Main Menu press ECG/RESP to display the
first ECG/RESP menu:
12. Press buttons as desired to adjust the display:
One of six V lead
electrode placement
sites for the C lead.
LL
LA
RA
RL
5-lead Configuration
V6
V1
Neonate 3-lead Configuration
RA
LA
LL
NET
OFF
HR
MON
MON
17:05:10
PRINT FAULT
II
BR
P
1
( 96)
BPM
mmHg
(M)
TEMP
Ld
2
100.4
98.6
1.8
97
121
79
S
D
P
2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
12
S
D
ECG
1
SIZE
ECG
1
LEAD
ECG
2
SIZE
RESP
SIZE
MORE
ADULT
ON NETWORK
T1
T2
ΔT
mV
cm
1
112
RESP
2
X
mV
cm
1
V
ALLEN KIM
ECG1 lead selected
ECG1 size selected
RESP waveform is displayed (with RESP
option)
RESP lead selected
RESP size selected
BR indicates respiration rate is derived from active CO
2
channel.
RR indicates CO
2
is not monitored and respiration rate is derived from ECG/RESP.
ECG2 lead
The ECG2 lead is always V, except when
the ECG1 lead is V (in that case the
ECG2 lead is II).