Switch from hardwired to wireless monitoring – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual

Page 80

background image

76

Acuity

Welch Allyn Propaq CS Vital Signs Monitor

Disconnect the hardwired monitor from the Acuity System

1.

To disconnect the Propaq CS monitor from the Acuity network, press the NET OFF
key on the upper left corner of the monitor display. (From the Main Menu you can also
press SETUP, ACUITY, NET OFF.)

2.

Within 15 seconds, disconnect the Acuity network cable from the Propaq CS monitor
side panel and the bedside jack. If the patient will no longer be monitored with this
monitor, turn off the monitor to erase trend information.

Switch from hardwired to wireless monitoring

1.

Press the NET OFF key on the upper left corner of the monitor display. (From the
Main Menu you can also press SETUP, ACUITY, NET OFF.)

2.

Disconnect the Acuity network cable from the Wireless Propaq CS Acuity connector.

The Wireless Propaq CS automatically seeks to establish a wireless connection to the
Acuity network.

Note

If the Propaq CS has the wireless option, be aware that after you disconnect the
hardwire connection and disconnect the Acuity network cable, the Wireless
Propaq CS will attempt to establish a wireless connection to Acuity as long as the
power is on.

ADULT

MON

17:05:10

PRINT FAULT

II

NET

OFF

ON NETWORK

mV

cm

1

Press to disconnect
from Acuity

Note

If you disconnect the Acuity network cable without first pressing NET OFF, this
will cause a momentary equipment alert at the monitor and Acuity.