Display or print trends, Display or print a single trend – Welch Allyn 246 Propaq CS Vital Signs Monitor v.3.7X - User Manual User Manual
Page 70

66
Printing & trends
Welch Allyn Propaq CS Vital Signs Monitor
Display or print trends
Display or print a single trend
1.
To display a patient data trend, press SETUP, TRENDS from the Main Menu. The
monitor displays the Trend Menu:
Trends are also displayed on the Main Menu if all waveforms except ECG1 are turned
off in the Wave Select Window.
2.
Press NEXT TREND as needed to display the desired trend.
3.
Press PRINT to print the displayed trend.
Note
The symbol indicates that the NIBP reading was taken in the presence of high
artifact while monitoring ECG with the Smartcuf motion artifact filter on. Artifact
can affect accuracy. To help reduce artifact, see
RESP TREND
TIME
MON
II
BR
P
1
( 96)
mmHg
(M)
TEMP
100.4
98.6
1.8
97
121
79
S
D
P
2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
S
D
PRINT PRINT
OXYCRG
T1
T2
ΔT
mV
cm
1
112
BPM
10:00
9:58
9:56
9:54
9:52
9:50
9:48
9:46
9:44
9:42
HH:MM
HR/PR SpO
2
RR/BR
BPM
%
Br/M
98
98
97
98
99
98
98
98
97
97
12
12
12
12
12
21
12
12
12
12
SRCH
OFF
98
94
97
97
98
98
ETCO
2
INCO
2
mmHg
38
38
38
38
38
38
38
38
38
38
mmHg
3
3
3
3
3
3
3
3
3
3
112
101
12
NEXT
TREND
Press to scroll
through the
recorded trends.
Press to print
the displayed
trend.
The monitor stores trends every
2 minutes (except for NIBP) for
up to 5 hours. After 5 hours,
newly recorded trends replace
the oldest trends.
NIBP trends are stored when
measured. Up to 128 NIBP
trends can be stored (up to 8
hours).
The TIME, HR/PR, and SpO2 are
stored for all trends.
OFF indicates the vital sign was
not being monitored.
ERR#x in an NIBP TREND
display or printout indicates an
NIBP equipment alert occurred.
page 57 for NIBP alert error
numbers and definitions.
Press to scroll to the next
vital sign trend (IBP, TEMP,
and NIBP).
Press to print
OxyCRG (see
page 69).
Numerics
that have
alarmed are
highlighted.
NIBP TREND
TIME
MON
BR
( 96)
mmHg
TEMP
100.4
98.6
1.8
97
79
P
2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
S
D
T1
T2
ΔT
10:01
9:46
9:31
9:20
9:08
8:52
8:47
8:41
8:35
8 30
HH:MM
HR/PR SpO
2
S D (M)
RR/BR
BPM
%
mmHg
Br/M
80
80
80
80
80
80
80
80
80
80
97
97
97
97
97
97
97
97
97
97
122 58 ( 85)
126 62 ( 86)
127 62 ( 86)
134 66 ( 89)
124 57 ( 86)
127 58 ( 88)
124 56 ( 85)
126 57 ( 84)
127 58 ( 85)
132 61 ( 89)
12
12
12
12
12
12
12
12
12
12
12