Campbell Scientific CR3000 Micrologger User Manual
Page 535

Appendix B. Status Table and Settings
535
Table 124. Status-Table Fields and Descriptions
Fieldname Description
Variable
Type
Default
Range
Edit?
Info
Type
MeasureTime
Time (μs) required to make 
the measurements in this 
scan, including integration 
and settling times. Processing 
occurs concurrent with this 
time so the sum of measure 
time and process time is not 
the time required in the scan 
instruction. This is a static 
value calculated at compile 
time. Assumes all 
measurement instructions 
will run each scan. 
Integer
Status
ProcessTime
Processing time (μs) of the 
last scan. Time is measured 
from the end of the EndScan 
instruction (after the 
measurement event is set) to 
the beginning of the EndScan 
(before the wait for the 
measurement event begins) 
for the subsequent scan. 
Calculated dynamically (on 
the fly). 
Integer
Status
MaxProcTime
Maximum time (μs) required 
to run through processing for 
the current scan. This value 
is reset when the scan exits. 
Calculated dynamically (on 
the fly). 
Integer
Yes 
Reset by 
changing to 0 
Status
BuffDepth
Shows the current Pipeline 
Mode processing buffer 
depth., which indicates how 
far processing is currently 
behind measurement. 
MaxBuffDepth
Gives the maximum number 
of buffers processing lagged 
measurement. 
LastSystemScan8
The last time the background 
calibration executed. 
Integer array
Status
LastSlowScan9
The last time SlowSequence 
scan(s) executed. 
Integer array
Status
SystemProcTime8,12
The time (μs) required to 
process the background 
calibration. 
Integer array
Status
SlowProcTime9,12
The time (μs) required to 
process SlowSequence 
scan(s). 
Integer array
Status
MaxSystemProcTime8,
13 
 
The maximum time (μs) 
required to process the 
background calibration. 
Integer array
Status