8 signal settling time, Figure 92: input voltage rise and transient decay, Table 58. ac noise rejection on large signals – Campbell Scientific CR3000 Micrologger User Manual
Page 291

Section 8. Operation
291
approximately 170 µs, leaving a maximum input-settling time of approximately
8160 µs (8333 µs - 170 µs). If the maximum input-settling time is exceeded, 60-
Hz line-noise rejection will not occur. For 50-Hz rejection, the maximum input
settling time is approximately 9830 µs (10,000 µs - 170 µs). The CR3000 does
not prevent or warn against setting the settling time beyond the half-cycle limit.
Table ac Noise Rejection on Large Signals
(p. 291)
lists details of the half-cycle ac-
power line-noise rejection technique.
Table 58. ac Noise Rejection on Large Signals
Applies to analog input voltage ranges mV1000 and mV5000.
ac Power Line
Frequency
Measurement
Integration
Time
CRBasic
Integration
Code
Default
Settling
Time
Maximum
Recommended
Settling Time*
60 Hz
250 μs x 2
_60Hz
3000 μs 8330
μs
50 Hz
250 μs x 2
_50Hz
3000 μs 10000
μs
*Excitation time and settling time are equal in measurements requiring excitation. The CR3000 cannot excite VX / EX or IX
excitation channels during A/D conversion. The one-half-cycle technique with excitation limits the length of recommended
excitation and settling time for the first measurement to one-half-cycle. The CR3000 does not prevent or warn against setting a
settling time beyond the one-half-cycle limit. For example, a settling time of up to 50000 µs can be programmed, but the CR3000
will execute the measurement as follows:
1. CR3000 turns excitation on, waits 50000 µs, and then makes the first measurement.
2. During A/D, CR3000 turns off excitation for ≈170 µs.
3. Excitation is switched on again for one-half cycle, then the second measurement is made.
Restated, when the CR3000 is programmed to use the half-cycle 50-Hz or 60-Hz rejection techniques, a sensor does not see a
continuous excitation of the length entered as the settling time before the second measurement if the settling time entered is greater
than one-half cycle. This causes a truncated second excitation. Depending on the sensor used, a truncated second excitation may
cause measurement errors.
8.1.2.8 Signal Settling Time
When the CR3000 switches to an analog input channel or activates excitation for
a bridge measurement, a settling time is required for the measured voltage to settle
to its true value before being measured. The rate at which the signal settles is
determined by the input settling time constant, which is a function of both the
source resistance and fixed input capacitance (3.3 nfd) of the CR3000.
Rise and decay waveforms are exponential. Figure Input Voltage Rise and
Transient Decay
(p. 291)
shows rising and decaying waveforms settling to the true
signal level, V
so
.
Figure 92: Input voltage rise and transient decay