Campbell Scientific CR3000 Micrologger User Manual
Page 308

Section 8. Operation
308
specification of 0.1°C for temperatures between 0 and 70°C. Below freezing and 
at higher temperatures, this specification is degraded. Combined with possible 
errors in the completion-resistor measurement and the Steinhart and Hart equation 
used to calculate the temperature from resistance, the accuracy of panel 
temperature is estimated in figure Panel Temperature Error Summary
(p. 309).
In
summary, error is estimated at ± 0.1°C over -0 to 40°C, ± 0.3°C from -25 to 50°C, 
and ± 0.8°C from -55 to 85°C. 
The error in the reference-temperature measurement is a combination of the error 
in the thermistor temperature and the difference in temperature between the panel 
thermistor and the terminals the thermocouple is connected to. The terminal strip 
cover should always be used when making thermocouple measurements. It 
insulates the terminals from drafts and rapid fluctuations in temperature as well as 
conducting heat to reduce temperature gradients. In a typical installation where 
the CR3000 is in a weather-tight enclosure not subject to violent swings in 
temperature or uneven solar radiation loading, the temperature difference between 
the terminals and the thermistor is likely to be less than 0.2°C. 
With an external driving gradient, the temperature gradients on the input panel 
can be much worse. For example, the CR3000 was placed in a controlled 
temperature chamber. Thermocouples in channels at the ends and middle of each 
analog terminal strip measured the temperature of an insulated aluminum bar 
outside the chamber. The temperature of this bar was also measured by another 
datalogger. Differences between the temperature measured by one of the 
thermocouples and the actual temperature of the bar are due to the temperature 
difference between the terminals the thermocouple is connected to and the 
thermistor reference (the figures have been corrected for thermistor errors). Figure 
Panel-Temperature Gradients (Low Temperature to High)
(p. 310)
shows the errors
when the chamber was changed from low temperature to high in approximately 15 
minutes. Figure Panel-Temperature Gradients (High Temperature to Low)
(p. 310)
shows the results when going from high temperature to low. During rapid 
temperature changes, the panel thermistor will tend to lag behind terminal 
temperature because it is mounted deeper in the CR3000. 
