beautypg.com

4 measurement accuracy, Timing – Campbell Scientific CR3000 Micrologger User Manual

Page 284

background image

Section 8. Operation

284 

 

 

Table 53. CRBasic Parameters Varying Measurement Sequence and

Timing

CRBasic Parameter

Description

MeasOfs

Correct ground offset on single-ended measurements.

RevDiff

Reverse high and low differential inputs.

SettlingTime

Sensor input settling time.

Integ

Duration of input signal integration.

RevEx

Reverse polarity of excitation voltage.

 

8.1.2.4 Measurement Accuracy

CR3000 analog-measurement error is calculated as

Error = Gain Error (%) + Offset Error 

Gain error is expressed as ±% of reading and is a function of CR3000
temperature. Between 0°C and 40°C, gain error is ±0.04% of input voltage. This
gain error assumes factory recalibration every two years.

Offset error depends on measurement type and input range. For differential
measurements with input reversal,

Offset Error = 1.5 • Resolution + 1.0 µV 

where

Basic Resolution is the published resolution of the programmed input 
voltage range (see CR3000 Specifications

 

). 

Figure CR3000 Voltage Measurement Accuracy

(p. 285)

illustrates that as magnitude

of input voltage decreases, measurement error decreases.

Note The accuracy specification includes only the CR3000 contribution to
measurement error. It does not include the error of sensors.

For example, assume the following (see CR3000 Specifications

):

• Input Voltage: 5000 mV
• Programmed Input Voltage Range: ±5000 mV (

mV5000

)

• Programmed Measurement Instruction: VoltDiff()
• Input Measurement Reversal:

True

• CR3000 Temperature: Between 0°C and 40°C

Accuracy of the measurement is calculated as follows:

Error = Gain Error + Offset Error, 

where

Gain Error = ± (5000 * 0.0004)