beautypg.com

1 ac power line noise rejection, Table 57. ac noise rejection on small signals – Campbell Scientific CR3000 Micrologger User Manual

Page 290

background image

Section 8. Operation

290 

 

 

8.1.2.7.1 ac Power Line Noise Rejection

Grid or mains power (50 or 60 Hz, 230 or 120 Vac) can induce electrical noise at
integer multiples of 50 or 60 Hz. Small analog voltage signals, such as
thermocouples and pyranometers, are particularly susceptible. CR3000 voltage
measurements can be programmed to reject (filter) 50-Hz or 60-Hz related noise.

ac Noise Rejection on Small Signals

The CR3000 rejects ac power line noise on all voltage ranges except

mV5000

and

mV1000

by integrating the measurement over exactly one ac cycle before A/D

conversion as illustrated in table ac Noise Rejection on Small Signals

(p. 290)

and

the full cycle technique of figure ac Power Line Noise Rejection Techniques

(p.

290).

Table 57. ac Noise Rejection on Small Signals

Applies to all analog input voltage ranges except

mV1000

and

mV5000

.

ac Power Line Frequency

Measurement Integration

Duration

CRBasic Integration Code

60 Hz

16.667 ms

_60Hz

50 Hz

20 ms

_50Hz

 

Figure 91: Ac power line noise rejection techniques

ac Noise Rejection on Large Signals

If rejecting ac-line noise when measuring with the 1000 mV (

mV1000

) and 5000

mV (

mV5000

) ranges, the CR3000 makes two fast measurements separated in time

by one-half line cycle (see figure ac Power Line Noise Rejection Techniques

(p. 290)

). A 60-Hz half cycle is 8333 µs, so the second measurement must start 8333 µs
after the first measurement integration began. The A/D conversion time is