beautypg.com

4 pulse measurement problems, 1 pay attention to specifications, 2 input filters and signal attenuation – Campbell Scientific CR3000 Micrologger User Manual

Page 325

background image

Section 8. Operation

325

 

Table 74. Frequency Resolution Comparison

0.5 s Scan

5.0 s Scan

PulseCount(),

POption

=

1

FR = 2 Hz

FR = 0.2 Hz

TimerIO(),

Function

=

2

FR = 0.0011 Hz

FR = 0.00011 Hz

 

Q — When more than one pulse is in a scan interval, what does TimerIO() return
when configured to return a frequency? Does it average the measured periods and
compute the frequency from that (f = 1/T)? For example:

Scan

(50,mSec,10,0)

TimerIO

(WindSpd(),11111111,00022000,60,Sec)

A — In the background, a 32-bit timer counter is saved each time the signal
transitions as programmed (rising or falling). This counter is running at a fixed
high frequency. A count is also incremented for each transition. When the
TimerIO() instruction executes, it uses the difference of time between the edge
prior to the last execution and the edge prior to this execution as the time
difference. The number of transitions that occur between these two times divided
by the time difference gives the calculated frequency. For multiple edges
occurring between execution intervals, this calculation does assume that the
frequency is not varying over the execution interval. The calculation returns the
average regardless of how the signal is changing.

8.1.5.4 Pulse Measurement Problems

8.1.5.4.1 Pay Attention to Specifications

The table Example of Differing Specifications for Pulse Input Channels

(p. 325)

compares specifications for pulse-input channels to emphasize the need for
matching the proper device to application. Take time to understand signals to be
measured and compatible channels.

Table 75. Example of Differing Specifications for Pulse-Input

Channels



Pulse Channels

P1, P2

Digital I/O Channels

C1, C2, C3, C4,

C5, C6, C7, C8

High Frequency Max

250 kHz

400 kHz

Max Input Voltage

20 Vdc

16 Vdc

State Transition

Thresholds

Count upon transition from

<0.9 to >2.2 Vdc

Count upon transition from

<1.2 to >3.8 Vdc

 

8.1.5.4.2 Input Filters and Signal Attenuation

Pulse-input channels are equipped with input filters to reduce spurious noise that
can cause false counts. The higher the time constant (τ) of the filter, the tighter
the filter. Table Time Constants

(p. 326)

lists τ values for pulse-input channels. So,

while TimerIO() frequency measurement may be superior for clean signals, a
pulse channel filter (much higher τ) may be required to get a measurement on a
dirty signal.