beautypg.com

1 input limits – Campbell Scientific CR3000 Micrologger User Manual

Page 281

background image

Section 8. Operation

281

 

instructions BrFull(), BrFull6W(), BrHalf4W(), TCDiff(), VoltDiff () and
Resistance () instructions instructions perform DIFF voltage measurements.

Figure 88: PGI amplifier

A PGIA processes the difference between the H and L inputs, while rejecting
voltages that are common to both inputs. Figure PGIA with Input Signal
Decomposition

(p. 281),

illustrates the PGIA with the input signal decomposed into a

common-mode voltage (V

cm

) and a DIFF-mode voltage (V

dm

). The common-mode

voltage is the average of the voltages on the V

H

and V

L

inputs, i.e., V

cm

= (V

H

+

V

L

)/2, which can be viewed as the voltage remaining on the H and L inputs with

the DIFF voltage (V

dm

) equal to 0. The total voltage on the H and L inputs is

given as V

H

= V

cm

+ V

dm

/2, and V

L

= V

cm

– V

dm

/2, respectively.

Figure 89: PGIA with input signal decomposition

8.1.2.1 Input Limits

The input limits specification is the voltage range, relative to CR3000 ground,
which both H and L input voltages must be within to be processed correctly by the
PGIA. Input limits for the CR3000 are ±5 Vdc. Input voltages in which V

H

or V

L

are beyond the ±5 Vdc input limits may suffer from undetected measurement
errors. The term “common-mode range”, which defines the valid range of
common-mode voltages, is often used instead of “input limits.” For DIFF voltages
that are small compared to the input limits, common-mode range is essentially
equivalent to input limits. Yet from figure PGIA with Input Signal Decomposition

(p. 281),

Common‐Mode Range = ± 5 Vdc – | V

dm

/2 |, 

indicating a reduction in common-mode range for increasing DIFF signal
amplitudes. For example, with a 5000 mV DIFF signal, the common-mode range