Section 7. Installation

161

 

Table 25. Calibration Report for Salinity Sensor

Parameter

Parameter at

Deployment

Parameter at 7-Day

Service

mV output

1350 mV

1345 mV

KnownSalt (standard
solution)

30 mg/l

30 mg/l

Multiplier

0.05 mg/l/mV

0.05 mg/l/mV

Offset

-37.50 mg/l

-37.23 mg/l

RH reading

30 mg/l

30 mg/l

1. Send the program in CRBasic example FieldCal Offset Demo Program

(p. 161)

to the CR3000. An excitation channel has been programmed to simulate a
sensor output.

2. To simulate the salinity sensor in deployment-calibration conditions (30 mg/l

standard solution), place a jumper wire between channels VX1/EX1 and SE6
(3L). Set variable mV to 1350. Set variable KnownSalt to 30.

3. To simulate the deployment calibration, change the value in variable CalMode

to 1 to start calibration. When CalMode increments to 6, offset calibration is
complete. Calibrated offset will equal -37.48 mg/l at this stage of the example.

4. To continue this example and simulate an offset-drift condition, change

variable mV to 1345.

5. To simulate 7-day, service-calibration conditions (30 mg/l standard solution),

set variable KnownSalt to 30.0. Change the value in variable CalMode to 1 to
start calibration. When CalMode increments to 6, 7-day, service-offset
calibration is complete. Calibrated offset will equal -37.23 mg/l.

CRBasic Example 27. FieldCal() Offset Demo Program 

'Jumper VX1 to SE6(3L) to simulate a sensor

Public

mV

'Excitation mV output

Public

KnownSalt

'Known salt concentration

Public

CalMode

'Calibration trigger

Public

Multiplier

'Multiplier (starts at .05 mg / liter / mV,

'does not change)

Public

Offset

'Offset (starts at zero, not changed)

Public

SaltContent

'Salt concentration

'Data Storage Output of Calibration Data -- stored whenever a calibration occurs

DataTable

(CalHist,NewFieldCal,200)

SampleFieldCal

EndTable