Section 7. Installation

160 

 

 

5. To simulate conditions for a 30-day, service-calibration, again with desiccated

chamber conditions, set variable KnownRH to 0.0. Change the value in
variable CalMode to 1 to start calibration. When CalMode increments to 6,
simulated 30-day, service zero calibration is complete. Calibrated Offset will
equal -52.5%.

CRBasic Example 26. FieldCal() Zeroing Demonstration Program 

'Jumper VX1 to SE6(3L) to simulate a sensor

Public

mV

'Excitation mV Output

Public

KnownRH

'Known Relative Humidity

Public

CalMode

'Calibration Trigger

Public

Multiplier

'Multiplier (Starts at .05 mg / liter / mV,

'does not change)

Public

Offset

'Offset (Starts at zero, not changed)

Public

RH

'Measured Relative Humidity

'Data Storage Output of Calibration Data -- stored whenever a calibration occurs

DataTable

(CalHist,NewFieldCal,200)

SampleFieldCal

EndTable

BeginProg

Multiplier = .05
Offset = 0

LoadFieldCal

(true)

'Load the CAL File, if possible

Scan

(100,mSec,0,0)

'Simulate measurement by exciting channel VX1/EX1

ExciteV

(Vx1,mV,0)

'Make the calibrated measurement

VoltSE

(RH,1,mV5000,6,1,0,250,Multiplier,Offset)

'Perform a calibration if CalMode = 1

FieldCal

(0,RH,1,Multiplier,Offset,CalMode,KnownRH,1,30)

'If there was a calibration, store it into a data table

CallTable

(CalHist)

NextScan

EndProg

 

7.8.1.5.2 Offset (Option 1)

Case: A sensor measures the salinity of water. Multiplier is known to be stable,
but sensor offset drifts and requires regular offset correction using a standard
solution. The following procedure offsets the measurement to obtain the
calibration report shown.