473

 

watchdog timer 

An error‐checking system that examines the processor state, software 
timers, and program‐related counters when the datalogger is running its 
program. If the processor has bombed or is neglecting standard system 
updates or if the counters are outside the limits, the watchdog timer 
resets the processor and program execution. Voltage surges and 
transients can cause the watchdog timer to reset the processor and 
program execution. When the watchdog timer resets the processor and 
program execution, an error count is incremented in the 
WatchdogTimer entry of the Status table

 (p. 530).

 A low number (1 to 10) 

of watchdog timer resets is of concern, but normally indicates the user 
should just monitor the situation. A large number (>10) of errors 
accumulating over a short period of time should cause increasing alarm 
since it indicates a hardware or software problem may exist. When large 
numbers of watchdog‐timer resets occur, consult with a Campbell 
Scientific applications engineer. 

weather tight 

Describes an instrumentation enclosure impenetrable by common 
environmental conditions. During extraordinary weather events, 
however, seals on the enclosure may be breached. 

Web API 

Application Programming Interface (see section Web API

,

 for more 

information). 

Glossary. Wild Card 

a character or expression that substitutes for any other character or 
expression. 

XML 

Extensible markup language. 

User Program 

The CRBasic program written by the CR3000 user in CRBasic Editor or 
Short Cut

11.2 Concepts

11.2.1 Accuracy, Precision, and Resolution

Three terms often confused are accuracy, precision, and resolution. Accuracy is a
measure of the correctness of a single measurement, or the group of
measurements in the aggregate. Precision is a measure of the repeatability of a