3 measurement rate: 601 to 2000 hz, Table 38. measuring voltse() at 2000 hz – Campbell Scientific CR3000 Micrologger User Manual
Page 240

Section 7. Installation
240
• One more way to view sub-scans is that they are a convenient (and only) way
to put a loop around a set of measurements. SubScan() / NextSubScan 
specifies a timed loop for so many times around a set of measurements that 
can be driven by the task sequencer. 
7.8.12.3 Measurement Rate: 601 to 2000 Hz
To measure at rates greater than 600 Hz, VoltSE() is switched into burst mode by 
placing a dash (
-
) before the channel number and placing alternate arguments in
other parameters. Alternate arguments are described in the table Parameters for 
Analog Burst Mode
(p. 241).
In burst mode, VoltSE() dwells on a single channel
and measures it at rates up to 2000 Hz, as demonstrated in the CRBasic example 
Measuring VoltSE() at 2000 Hz
.
The example program has an 86% duty cycle.
That is, it makes measurements over only the leading 86% of the scan. Note that 
burst mode places all measurements for a given burst in a single variable array 
and stores the array in a single (but very long!) record in the data table. The exact 
sampling interval is calculated as, 
Tsample = 1.085069 * INT((SettleUSEC / 1.085069) + 0.5
where
SettleUSEC
is the sample interval (µs) entered in the
SettlingTime
parameter
of the analog input instruction.
Table 38. Measuring VoltSE() at 2000 Hz
PipeLineMode
'<<<
Public
BurstSE(1735)
DataTable
(BurstSETable,1,-1)
Sample
(1735,BurstSE(),FP2)
EndTable
BeginProg
Scan
(1,Sec,10,0)
'
Measurement speed and count are set within VoltSE()
VoltSe
(BurstSE(),1735,mV2_5,-1,False,500,0,1.0,0)
CallTable
BurstSETable
NextScan
EndProg
Many variations of the burst program are possible. Multiple channels can be 
measured, but one channel burst is completed before the next begins. Caution 
dictates that a specific configuration be thoroughly tested before deployment. 
