beautypg.com

Campbell Scientific CR3000 Micrologger User Manual

Page 326

background image

Section 8. Operation

326 

 

 

Input filters, however, attenuate the amplitude (voltage) of the signal. The
amount of attenuation is a function of the frequency passing through the filter.
Higher-frequency signals are attenuated more. If a signal is attenuated enough, it
may not pass the state transition thresholds required by the detection device (listed
in table Pulse-Input Channels and Measurements

(p. 40)

). To avoid over

attenuation, sensor output voltage must be increased at higher frequencies. As an
example, table Filter Attenuation of Frequency Signals

(p. 326)

lists low-level ac

frequencies and the voltages required to overcome filter attenuation.

For pulse-input channels P1P4, an RC input filter with an approximate 1.2-

μs

time constant precedes the inverting CMOS input buffer. The resulting amplitude
reduction is illustrated in figure Amplitude Reduction of Pulse-Count Waveform

(p.

327).

For a 0- to 5-Vdc square wave applied to a pulse channel, the maximum

frequency that can be counted in high-frequency mode is approximately 250 kHz.

Table 76. Time Constants (τ)

Measurement

τ

Pulse channel, high-frequency mode

1.2

Pulse channel, switch-closure mode

3300

Pulse channel, low-level ac mode

See table Filter Attenuation of

Frequency Signals

(p. 326)

footnote

Digital I/O, high-frequency mode

0.025

Digital I/O, switch-closure mode

0.025

 

Table 77. Filter Attenuation of Frequency Signals.

As shown for low-level ac inputs, increasing voltage is required at increasing frequencies to
overcome filter attenuation on pulse-input channels*.

ac mV (RMS)

Maximum Frequency

20

200

2000
5000

20

200

10,000
20,000

*8.5-ms time constant filter (19 Hz 3 dB frequency) for low-amplitude signals. 1-ms time
constant (159 Hz 3 dB frequency) for larger (> 0.7 V) amplitude signals.