beautypg.com

1 input and excitation reversal – Campbell Scientific CR3000 Micrologger User Manual

Page 288

background image

Section 8. Operation

288 

 

 

Table 55. Analog Measurements and Offset Voltage Compensation


CRBasic

Voltage

Measurement

Instruction


Input

Reversal

(

RevDiff

=

True

)


Excitation

Reversal

(

RevEx

=

True

)

Measure

Ground

Reference

Offset

(

MeasOff

=

True

)

Background

Calibration

(RevDiff = False)

(RevEx = False)

(MeasOff = False)

VoltDiff()

*

*

VoltSe()

* *

TCDiff()

*

*

TCSe()

* *

BrHalf()

* *

BrHalf3W()

* *

Therm107()

* *

Therm108()

* *

Therm109()

* *

BrHalf4W()

* * *

BrFull()

* * *

BrFull6W()

* * *

AM25T()

* * *

 

8.1.2.6.1 Input and Excitation Reversal

Reversing inputs (differential measurements) or reversing polarity of excitation
voltage (bridge measurements) cancels stray voltage offsets. For example, if there
is a 3-

μV offset in the measurement circuitry, a 5-mV signal is measured as 5.003

mV. When the input or excitation is reversed, the measurement is -4.997 mV.
Subtracting the second measurement from the first and dividing by two cancels
the offset:

5.003 mV - (-4.997 mV) = 10.000 mV
10.000 mV / 2 = 5.000 mV.

When the CR3000 reverses differential inputs or excitation polarity, it delays the
same settling time after the reversal as it does before the first measurement. So,
there are two delays per channel when either

RevDiff

or

RevEx

is used. If both

RevDiff

and

RevEx

are

True

, four measurements are performed; positive and

negative excitations with the inputs one way and positive and negative excitations
with the inputs reversed. To illustrate,

1. the CR3000 switches to the channel

2. sets the excitation, settles, measures,

3. reverses the excitation, settles, measures,

4. reverses the excitation, reverses the inputs, settles, measures,

5. reverses the excitation, settles, measures.

There are four delays per channel measured. The CR3000 processes the four sub-
measurements into a single reported value. In cases of excitation reversal,