74

Note

The leakage current range depends on the measuring
network used. The table below shows the maximum
and minimum values of each range for each network
type.

MD A~F,I

AC, DC,
AC+DC

Range

25.00mA

5.000mA

500.0uA

50.00uA

Maximum 25.00mA 5.000mA 500.0uA

50.00uA

Minimum 4.00mA

0.400mA

40.00uA

4.00uA

AC Peak

Range

75.0mA

10.00mA

1.000mA

500.0uA

Maximum 75.0mA

10.00mA

1.000mA

500.0uA

Minimum 8.0mA

0.80mA

0.100mA

40.0uA

MD G

AC, DC,
AC+DC

Range

25.00mA

5.000mA

500.0uA

50.00uA

Maximum 16.00mA 3.300mA 330.0uA

33.00uA

Minimum 3.00mA

0.300mA

30.00uA

4.00uA

AC Peak

Range

75.0mA

10.00mA

1.000mA

500.0uA

Maximum 50.0mA

6.60mA

0.660mA

330.0uA

Minimum 6.0mA

0.60mA

0.070mA

30.0uA

MD H

AC, DC,
AC+DC

Range

25.00mA

5.000mA

500.0uA

50.00uA

Maximum 12.50mA 2.500mA 250.0uA

25.00uA

Minimum 2.00mA

0.200mA

20.0uA

4.00uA

AC Peak

Range

75.0mA

10.00mA

1.000mA

500.0uA

Maximum 37.5mA

5.00mA

0.500mA

250.0uA

Minimum 4.0mA

0.40mA

0.050mA

20.0uA