Patient Auxiliary Current

Network

Medical network

Connection

As illustrated below, ensure that the power source
and test leads are properly connected.

Medical Network

1. Connect the test leads to the T1 and T2

terminals.

2. Position the test leads to the applied part of the

EUT.

Measurement
Setup

Confirm all settings including MD, leakage current
mode, measurement time, upper and lower limits,
and other parameters.

Panel Operation

1. Turn on the circuit breaker.

Connect the EUT power
terminal.



GLC-9000

2. Press the START button to

start measurements.



GLC-9000