beautypg.com

Measurement network (md) – GW Instek GLC-9000 User Manual User Manual

Page 179

background image

APPENDIX

179

Measurement Network (MD)

MD

Circuit

R.C. parameters*

Standards

Compliance

A

Rs:1.5kΩ
Rb:0.5kΩ

Cs:0.22uF

GB/T12113
IEC60990
GB4943
IEC60995
UL3101
JIS C1010-1

B

Rs:1.5kΩ
Rb:0.5kΩ
R1:10kΩ

Cs:0.22uF
C1:0.022uF

GB/T12113
IEC60990
GB4943
IEC60995
UL3101
JIS C1010-1

C

Rs:1.5kΩ
Rb:0.5kΩ
R2:10kΩ
R3:20kΩ

Cs:0.22uF
C2:6.2nF
C3:9.1nF

GB/T12113
IEC60990
GB4943
IEC60995
UL3101
JIS C1010-1

D

Rb:0.5kΩ C1:0.45uF

IEC60335-1
UL1563

E

Rb:1kΩ

UL3101
JIS C9250-92

F

Rb:1kΩ
R1:10kΩ

C1:15nF IEC60601-1

UL2601-1
EN60601-1
UL3111
JIS T1001-92